This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Various Micro Therapeutics Inc. catheters and devices (2016-10-18)

Starting date:
October 18, 2016
Posting date:
November 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60914

Affected Products

  1. ALLIGATOR RETRIEVAL DEVICE
  2. X-CELERATOR EXCHANGE HYDROPHILIC GUIDEWIRE - SS
  3. ULTRAFLOW HPC FLOW DIRECTED MICRO CATHETER
  4. MARATHON FLOW DIRECTED MICRO CATHETER
  5. PIPELINE EMBOLIZATION DEVICE (25-35MM)

Reason

Medtronic has identified the potential for an issue with a specific set of lot numbers where the PTFE (polytetrafluoroethylene) coating could delaminate and separate from the delivery wire. Delamination and detachment of the PTFE coating material may lead to PTFE coating in the blood stream. PTFE in the blood stream, based on its specific size and quantity, may be safe or could lead to thromboembolic complications or irreversible injuries including but not limited to: intracranial edema, peripheral edema, incomplete treatment, infection, local inflammatory response, systemic inflammatory response, ischemic stroke, lysis/necrosis, neurological deficit, organ impairment, shock, space occupying lesion, thrombosis, and hemorrhage.



Through September 27, 2016, Medtronic has received a total of five (5) reports for the products that are potentially affected by this issue. Medtronic has received no reports of serious injuries or patient death as a result of this issue.

Affected products

A. ALLIGATOR RETRIEVAL DEVICE

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • FA-88810-20
  • FA-88810-30
  • FA-88810-40
  • FA-88810-50
Companies
Manufacturer

Micro Therapeutics Inc.

DBA EV3 Neurovascular

9775 Toledo Way

Irvine

92618

UNITED STATES



B. X-CELERATOR EXCHANGE HYDROPHILIC GUIDEWIRE - SS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 103-0601-300
  • 103-0602-300
Companies
Manufacturer

Micro Therapeutics Inc.

DBA EV3 Neurovascular

9775 Toledo Way

Irvine

92618

UNITED STATES



C. ULTRAFLOW HPC FLOW DIRECTED MICRO CATHETER

Lot or serial number
  • A072386
  • A092709
  • A126396
  • A127019
  • A091993
  • A091996
  • 9972750
  • A020842
  • A025090
  • A044668
Model or catalog number
  • 105-5065
Companies
Manufacturer

Micro Therapeutics Inc.

DBA EV3 Neurovascular

9775 Toledo Way

Irvine

92618

UNITED STATES



D. MARATHON FLOW DIRECTED MICRO CATHETER

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 105-5055
Companies
Manufacturer

Micro Therapeutics Inc.

DBA EV3 Neurovascular

9775 Toledo Way

Irvine

92618

UNITED STATES



E. PIPELINE EMBOLIZATION DEVICE (25-35MM)

Lot or serial number
  • A090167
  • A090181
  • A090209
  • A104383
  • A104400
  • A104409
  • A110383
  • A110389
Model or catalog number
  • FA-71375-25
  • FA-71400-25
  • FA-71425-25
  • FA-71450-25
  • FA-71475-25
  • FA-71500-25
Companies
Manufacturer

Micro Therapeutics Inc.

DBA EV3 Neurovascular

9775 Toledo Way

Irvine

92618

UNITED STATES