Health product recall

Various ID-MTS Gel Cards products (2018-04-25)

Starting date:
April 25, 2018
Posting date:
May 4, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66704

Affected products

A. MTS Buffered Gel Card

B. MTS Anti-IgG Card

C. MTS A/B/D Monoclonal Grouping Card

D. MTS A/B/D/Monoclonal and Reverse Grouping Card

E. MTS IgG,-C3d Card

F. MTS Monoclonal Control Card

G. MTS Anti-A,B Card

H. MTS Anti-E Card

I. MTS Anti-C Card

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall as it may potentially affect the ID-MTS Gel Cards. The gel card trays are stored in paper board trays. The gel card trays may contain an increased amount of white dust/debris from the trays that can adhere to the outer surface of the cards. Ortho has confirmed reports of white dust/debris on the outer surface of various ID-MTS Gel Cards products. While the issue does not affect use of the ID-MTS Gel Cards in manual method, Ortho has identified that the adhering of white dust/debris on the outer surface of the cards may potentially affect the performance of the cards on ORTHO VISION and ORTHO VISION Max Analyzer.

Affected products

A. MTS Buffered Gel Card

Lot or serial number

011518004-01

092917004-01

101217004-01

120417004-01

Model or catalog number

MTS085014

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



B. MTS Anti-IgG Card

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

MTS084024

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



C. MTS A/B/D Monoclonal Grouping Card

Lot or serial number

010818053-01

010818053-02

010818053-03

010818053-04

081117053-02

081117053-03

081117053-04

101317053-02

101317053-03

101317053-04

Model or catalog number

MTS081115

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



D. MTS A/B/D/Monoclonal and Reverse Grouping Card

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

MTS080515

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



E. MTS IgG,-C3d Card

Lot or serial number

062617005-01

100417005-01

111317005-01

Model or catalog number

MTS084014

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



F. MTS Monoclonal Control Card

Lot or serial number

102417042-01

Model or catalog number

MTS081001

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



G. MTS Anti-A,B Card

Lot or serial number

083017038-01

121317038-01

Model or catalog number

MTS080016

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



H. MTS Anti-E Card

Lot or serial number

102617047-01

Model or catalog number

MTS080212

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES



I. MTS Anti-C Card

Lot or serial number

110217045-01

Model or catalog number

MTS080210

Companies
Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

UNITED STATES