This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Various Devices of Mandible External Fixator System (2014-10-09)

Starting date:
October 9, 2014
Posting date:
November 7, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42065

Recalled Products

A) Mandible External Fixator System - Adjustable Clamp

B) Mandible external fixator system - Schanz Screw

C) Mandible external fixator system - Mandible Rod

Reason

Certain components within the CMF Mandible External Fixator System have been labelled and/or etched with information indicating they are "MR safe". Changes have been made to testing standards (ASTM  F2503) to designate a product safe and compatible in the MR environment.  Per current ASTM  F2503 testing standards, metal devices are not to be identified as MR safe.

Affected products

A) Mandible External Fixator System - Adjustable Clamp

Lot or serial number

All lots.

Model or catalog number

03.305.006

03.305.105

Companies
Manufacturer

Synthes (Canada) Ltd.

2566 Meadowpine Blvd.

Mississauga

L5N 6P9

Ontario

CANADA



B) Mandible external fixator system - Schanz Screw

 

Lot or serial number

All lots.

Model or catalog number

04.305.003

04.305.004

04.305.005

04.305.010

04.305.011

04.305.012

04.305.128

04.305.130

04.305.134

04.305.138

04.305.140

04.305.142

 

Companies
Manufacturer

Synthes (Canada) Ltd.

2566 Meadowpine Blvd.

Mississauga

L5N 6P9

Ontario

CANADA



C) Mandible external fixator system - Mandible Rod

Lot or serial number

All lots.

Model or catalog number

04.305.100

04.305.101

04.305.102

04.305.103

Companies
Manufacturer

Synthes (Canada) Ltd.

2566 Meadowpine Blvd.

Mississauga

L5N 6P9

Ontario

CANADA