This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

VANCOMYCIN ASSAYS (2015-09-04)

Starting date:
September 4, 2015
Posting date:
October 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55212

Affected Products

  1. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 701/702 SYSTEMS
  2. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 501/502 SYSTEMS
  3. VANCOMYCIN (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI 912 AND MODULAR P ANALYZERS)

Reason

An inaccurately low result was reported for a Vancomycin patient sample on Cobas C501. This was confirmed by Roche. Detailed investigations revealed that the sample contains a component which inhibits the drug labeled enzyme glucose-6-phosphate dehydrogenase (Vanc-G6PDH) in the emit assay.

Affected products

A. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 701/702 SYSTEMS

Lot or serial number

All lots.

Model or catalog number
  • 5842310190
Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



B. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 501/502 SYSTEMS

Lot or serial number

All lots.

Model or catalog number
  • 4491050190
Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



C. VANCOMYCIN (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI 912 AND MODULAR P ANALYZERS)

Lot or serial number

All lots.

Model or catalog number
  • 4642481190
  • 4642490190
Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY