Health product recall

V-Loc™ 90 and 180 Suture Products

Brand(s)
Last updated

Summary

Product
V-Loc™ 90 and 180 Suture Products
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

V-Loc 180 Wound Closure Device Needle P-12

A6A0972VFY

VLOCL0014

V-Loc 180 Abs Close Device GS-21

A5M1060VY

VLOCL0316

V-Loc 90 Absorbable Wound Closure Device P-12

A5M1022VFY

VLOCM0003

V-Loc 180 Absorbable Wound Closure Device - Green Color

A6A0998VY

VLOCL2246

V-Loc 180 Absorbable Wound Closure Device - Green Color

A6A0996VY

VLOCL0306

V-Loc 180 Wound Closure Device Needle V-20

A5M1180VY

VLOCL0604

V-Loc 180 Wound Closure Device Needle V-20

A5M1181VY

VLOCL0604

V-Loc 180 Wound Closure Device Needle CV-23

A5M1064VY

VLOCL0804

Issue

On 29-Jan-2026 during a quality inspection at a manufacturing site, a V-Loc 180 unit was found with a hole in the breather pouch on the Tyvek side. The breather pouch serves as the sterile barrier system for the suture. Investigation determined the briller plate on the sealing station may be causing the hole. No complaints have been identified for this issue, but lots within the issue bounding have been released to Distribution Centers (DCs) and sold to customers. These released lots may also contain units with holes in the Tyvek side of the breather pouch.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Covidien Llc

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82349

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe