UniMatch Plus Software (2019-01-31)
- Starting date:
- January 31, 2019
- Posting date:
- February 15, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69048
Affected Products
UniMatch Plus Software
Reason
The UniMatch Plus Software that uses lot specific software catalog files (CHR_005_201807v1.uch) contain incorrect specificity information for lane 63, which may result in an ambiguous result or a mistyping. The impacted software catalog files were released and available for use between the date period Oct 12 2018 and Nov 15, 2018. Only files analyzed during this period are impacted. Worksheets for the accessory product are correct.
Affected products
UniMatch Plus Software
Lot or serial number
CHR_005_201807v1.uch
Model or catalog number
A10286
Companies
- Manufacturer
-
One Lambda Inc
21001 Kittridge Street
Canoga Park
91303-2801
California
UNITED STATES