This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

UNIFY, FORTIFY AND QUADRA (2016-10-07)

Starting date:
October 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60698

Affected Products

A. UNIFY

B. FORTIFY ST VR

C. FORTIFY ST DR

D. FORTIFY ASSURA DR

E. FORTIFY ASSURA VR

F. QUADRA ASSURA MP

Reason

St. Jude Medical has identified a risk of premature battery depletion associated with certain St. Jude medical implantable cardioverter defibrillator (ICD) and cardiac resynchronisation therapy defibrillator (CRT-D) devices manufactured before May 23, 2015.  There is evidence of lithium material in the form of "clusters" in the battery. Some devices exhibited visible clusters bridging the cathode and anode causing shorting. Lithium cluster formation is a known phenomenon with this type of battery.

Affected products

A. UNIFY

Lot or serial number

Not Applicable

Model or catalog number

CD3231-40

CD3235-40

CD3235-40Q

Companies
Manufacturer

St. Jude Medical, Cardiac Rhythm Management Division

15900 VALLEY VIEW COURT

Sylmar

91342

California

UNITED STATES



B. FORTIFY ST VR

Lot or serial number

Not Applicable

Model or catalog number

CD1235-40

CD1235-40Q

Companies
Manufacturer

St. Jude Medical, Cardiac Rhythm Management Division

15900 VALLEY VIEW COURT

Sylmar

91342

California

UNITED STATES



C. FORTIFY ST DR

Lot or serial number

Not Applicable

Model or catalog number

CD2235-40

CD2235-40Q

Companies
Manufacturer

St. Jude Medical, Cardiac Rhythm Management Division

15900 VALLEY VIEW COURT

Sylmar

91342

California

UNITED STATES



D. FORTIFY ASSURA DR

Lot or serial number

Not Applicable

Model or catalog number

CD2259-40

CD2259-40Q

CD2359-40C

CD2359-40QC

Companies
Manufacturer

St. Jude Medical, Cardiac Rhythm Management Division

15900 VALLEY VIEW COURT

Sylmar

91342

California

UNITED STATES



E. FORTIFY ASSURA VR

Lot or serial number

Not Applicable 

Model or catalog number

CD1259-40

CD1259-40Q

CD1359-40C

CD1359-40QC

Companies
Manufacturer

St. Jude Medical, Cardiac Rhythm Management Division

15900 VALLEY VIEW COURT

Sylmar

91342

California

UNITED STATES



F. QUADRA ASSURA MP

Lot or serial number

Not Applicable

Companies
Manufacturer

St. Jude Medical, Cardiac Rhythm Management Division

15900 VALLEY VIEW COURT

Sylmar

91342

California

UNITED STATES