Health product recall

Unicel DXI 600 and DXI 800 Access Immunoassay Systems (2014-01-29)

Starting date:
January 29, 2014
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38327

Recalled Products

Unicel DXI 600 and DXI 800 Access Immunoassay Systems

Reason

The Access Thyroglobulin Calibrators are a lyophilized product and as such need to be reconstituted prior to use. The reconstitution time of 30 minutes as stated in the thyroglobulin calibrators instructions for use (IFU) is insufficient for the four affected lots of thyroglobulin calibrators. This has resulted in high patient and control results that are on average greater than 1 standard deviation high.

Affected products

A. Unicel DXI 600 Access Immunoassay System - Thyroglobulin Assay

Lot or serial number

324753

325451

327168

330203

Model or catalog number

33865

Companies
Manufacturer

Beckman Coulter, Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



B. Access System - Thyroglobulin Assay

Lot or serial number

324753

325451

327168

330203

Model or catalog number

33865

Companies
Manufacturer

Beckman Coulter, Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



C. Unicel DXI 800 Access Immunoassay System - Thyroglobulin Assay

Lot or serial number

324753

325451

327168

330203

Model or catalog number

33865

Companies
Manufacturer

Beckman Coulter, Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES