Unicel DXI 600 and DXI 800 Access Immunoassay Systems (2014-01-29)
- Starting date:
- January 29, 2014
- Posting date:
- March 18, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-38327
Recalled Products
Unicel DXI 600 and DXI 800 Access Immunoassay Systems
Reason
The Access Thyroglobulin Calibrators are a lyophilized product and as such need to be reconstituted prior to use. The reconstitution time of 30 minutes as stated in the thyroglobulin calibrators instructions for use (IFU) is insufficient for the four affected lots of thyroglobulin calibrators. This has resulted in high patient and control results that are on average greater than 1 standard deviation high.
Affected products
A. Unicel DXI 600 Access Immunoassay System - Thyroglobulin Assay
Lot or serial number
324753
325451
327168
330203
Model or catalog number
33865
Companies
- Manufacturer
-
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. Access System - Thyroglobulin Assay
Lot or serial number
324753
325451
327168
330203
Model or catalog number
33865
Companies
- Manufacturer
-
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
C. Unicel DXI 800 Access Immunoassay System - Thyroglobulin Assay
Lot or serial number
324753
325451
327168
330203
Model or catalog number
33865
Companies
- Manufacturer
-
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES