UniCel DxI 600 Access Immunoassay System-Total T4 Assay
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
UniCel DxI 600 Access Immunoassay System-Total T4 Assay |
538815 |
33805 |
UniCel DxI 600 Access Immunoassay System-Total T4 Assay |
539243 |
33805 |
UniCel DxI 600 Access Immunoassay System-Total T4 Assay |
538729 |
33805 |
UniCel DxI 600 Access Immunoassay System-Total T4 Assay |
538898 |
33805 |
UniCel DxI 600 Access Immunoassay System-Total T4 Assay |
539107 |
33805 |
Issue
Access Total T4 calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. In the worst-case scenario, there is a medium safety risk associated with this issue due to the potential to cause false low patient results.
Additional information
Details
| Beckman Coulter, Inc. |
| 250 S. Kraemer Blvd., Brea, California, United States, 92821 |