Unicel DXH800 Coulter Cellular Analysis System – Biohazard or chemical exposure
- Starting date:
- April 3, 2012
- Posting date:
- May 14, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-26295
Recalled Products
- Unicel DXH800 Coulter Cellular Analysis System
Reason
The NRBC, Differential or Reticulocyte Mix Chamber Drain ports may become plugged by rubber pieces from specimen tube caps, and result in biohazard or chemical exposure and/or the potential for generating incorrect results.
Affected products
A. Unicel DXH800 Coulter Cellular Analysis System
Lot or serial number
- All Lots
Model or catalog number
- 629029
Companies
- Manufacturer
-
Beckman Coulter, Inc.