Health product recall

UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM (2017-11-06)

Starting date:
November 6, 2017
Posting date:
November 24, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65164

Affected products

UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM

Reason

The shield for the single-tube station on the DxH instrument may interfere with sample aspiration in the open vial position. The mechanical interference between the shield and the right (green) position of the single-tube station can cause the tube holder to remain in the down position during aspiration and then spring up, resulting in the dislodging of an open vial. This mechanical interference may result in a biohazardous spill. The left (lavender) position tube holder of the manual station is not impacted by this issue.

Affected products

UNICEL DXH 800 CELLULAR ANALYSIS SYSTEM

Lot or serial number

SERIAL # AS23151
SW VERSION 3.1.1.0

Model or catalog number

629029
B24802

Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES