Health product recall

Unicel Dxh 800 and 600 Coulter Cellular Analysis System (2017-04-25)

Starting date:
April 25, 2017
Posting date:
May 16, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63312

Affected Products

A. Unicel Dxh 800 Coulter Cellular Analysis System

B. Unicel Dxh 600 Coulter Cellular Analysis System

Reason

There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.

Affected products

A. Unicel Dxh 800 Coulter Cellular Analysis System

Lot or serial number

All lots

Model or catalog number

629029

B24802

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



B. Unicel Dxh 600 Coulter Cellular Analysis System

Lot or serial number

All lots

Model or catalog number

B23858

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES