Health product recall

Unicel DXH 600 and 800 Systems (2013-10-31)

Starting date:
October 31, 2013
Posting date:
December 17, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37193

Recalled products

  1. Unicel DXH 800 Coulter Cellular Analysis System
  2. Unicel DXH 600 Coulter Cellular Analysis System

Reason

Sample analysis cycle time for specimen with high white blood cell (WBC) counts (approximately 45.0 x 103 /uL or higher) is approximately two (2) minutes longer than expected for the DXH analytical systems with the software revisions Unicel DXH 800 Coulter cellular analysis system, software revision 2.0.2.1 and unicel DXH 600 Coulter cellular analysis system, software revision 1.0. The original total sample analysis cycle time (analysis and cleaning cycle) was approximately two and a half (2 1/2) minutes while the current cycle time is approximately five and a half (5 1/2) minutes. Test results are not affected by this issue.

Affected products

A. Unicel DXH 800 Coulter Cellular Analysis System

 

Lot or serial number

All lots

Model or catalog number
  • 629029
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



B. Unicel DXH 600 Coulter Cellular Analysis System

Lot or serial number

All lots

Model or catalog number
  • B23858
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES