Unicel DXH 600 and 800 Systems (2013-10-31)
- Starting date:
- October 31, 2013
- Posting date:
- December 17, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-37193
Recalled products
- Unicel DXH 800 Coulter Cellular Analysis System
- Unicel DXH 600 Coulter Cellular Analysis System
Reason
Sample analysis cycle time for specimen with high white blood cell (WBC) counts (approximately 45.0 x 103 /uL or higher) is approximately two (2) minutes longer than expected for the DXH analytical systems with the software revisions Unicel DXH 800 Coulter cellular analysis system, software revision 2.0.2.1 and unicel DXH 600 Coulter cellular analysis system, software revision 1.0. The original total sample analysis cycle time (analysis and cleaning cycle) was approximately two and a half (2 1/2) minutes while the current cycle time is approximately five and a half (5 1/2) minutes. Test results are not affected by this issue.
Affected products
A. Unicel DXH 800 Coulter Cellular Analysis System
Lot or serial number
All lots
Model or catalog number
- 629029
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. Unicel DXH 600 Coulter Cellular Analysis System
Lot or serial number
All lots
Model or catalog number
- B23858
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES