Ultraview SL™ Command Module
Brand(s)
Last updated
Summary
Product
Ultraview SL™ Command Module
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Ultraview SL™ Command Module |
More than 10 numbers, contact manufacturer. |
91496 |
Issue
It was discovered during production testing that the cardiac output function of the multiparameter module model 91496 was intermittently yielding bad or irregular curves.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Spacelabs Healthcare Inc.
35301 Se Center Street, Snoqualmie, Washington, United States, 98065
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81820
Get notified