Displaying 1 - 2 of 2 items.
Software versions prior to 2.04 show a waveform while in suspend processing, while software version 2.04 and above do not show a waveform. Both software versions show an error message stating "ECG processing suspended" in the waveform zone, "?" in the…
RecallHealth product recall | 2023-06-02
An error associated with the operating system causes this defect. After approximately 25 days of continuous use, technical alarm escalation does not occur as specified and expected. All clinical alarms functions as intended and…
RecallHealth product recall | 2022-06-13