Ultraview SL2600,SL Compact Monitor and SL Bedside/Central Monitor (2014-09-18)
- Starting date:
- September 18, 2014
- Posting date:
- September 22, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41437
Recalled products
- Ultraview SL2600 Compact Monitor
- Ultraview SL Compact Monitor
- Ultraview SL Bedside/Central Monitor
Reason
Potential hazard associated with the use of all patient monitors equipped with the perioperative option (-D). Operating room staff may not be aware that alarm tones are off if the device has been left in the end case state. This is more likely to occur in emergency surgery situations and with surgical teams that are not familiar with the monitor's operation.
Affected products
A. Ultraview SL2600 Compact Monitor
Lot or serial number
- 1370-004133
Model or catalog number
- 91370
Companies
- Manufacturer
-
Spacelabs Healthcare Inc.
35301 SE Center Street
Snoqualmie
98065
Washington
UNITED STATES
B. Ultraview SL Compact Monitor
Lot or serial number
- 1369-007338
Model or catalog number
- 91369
Companies
- Manufacturer
-
Spacelabs Healthcare Inc.
35301 SE Center Street
Snoqualmie
98065
Washington
UNITED STATES
C. Ultraview SL Bedside/Central Monitor
Lot or serial number
- 1387-005006
- 1387-102371
- 1387-102372
- 1387-102373
Model or catalog number
- 91387
Companies
- Manufacturer
-
Spacelabs Healthcare Inc.
35301 SE Center Street
Snoqualmie
98065
Washington
UNITED STATES