This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Ultrasound Gastroscopes (2014-06-28)

Starting date:
June 28, 2014
Posting date:
July 10, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40401

Recalled Products

Ultrasound Gastroscopes

Reason

Pentax medical has been notified of a potential risk related to the use of certain models of ultrasound gastro scope listed above in combination with aspiration needles if the instructions for use document are not followed carefully by the user. Specifically, if the user fails to properly position the elevator while advancing the aspiration needles through the endoscope, then the needle can derail from the elevator. This could result in the needles not being visible in the endoscopic or sono- graphic field of view, leaving to the possibility of patient injury such as perforation.

Affected products

Ultrasound Gastroscopes

Lot or serial number

A01023

A110468

A110032

A110469

A110072

E01041

A110079

E01052

A110157

E110164

A110158

E110175

A110159

E110190

A110176

E110201

A110185

E110206

A110188

E110207

A110190

E110229

A110191

E110243

A110197

E110312

A110198

E110332

A110201

E110388

A110202

E110392

A110220

E110395

A110223

E110400

A110247

A110260

A110275

A110278

A110308

A110312

A110317

A110319

A110327

A110328

A110329

A110330

A110331

A110362

A110365

A110428

A110430

A110446

A110456

Model or catalog number

EG-3630U

EG-3830UT

EG-3870UTK

FG-36UX

Companies
Manufacturer

Hoya Corporation

2-7-5 Naka-Ochiai, Shinjuku-ku,

Tokyo, Tokyo-to

161-8525

JAPAN