Health product recall

Twin Coil

Starting date:
May 9, 2017
Posting date:
June 5, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63486

Reason

The coil housing broke apart due to a breakdown in the coil winding element inside the housing. The breakdown occurred near the outlet from the outer winding. The breakdown had been initiated from mechanical stress near the outlet. Such breakdown can start in the wires that are bended or twisted. During stimulation some mechanical displacement/movements will occur inside the litze wire and between the windings.

Affected products

Twin Coil

Lot or serial number

Serial number: 1023

Model or catalog number

9016E0691

Companies

Manufacturer

Tonica Elektronik A/S

Lucernemarken 15

Farum

3520

DENMARK