Health product recall

TRILOGY (100) VENTILATOR SYSTEMS (2016-01-14)

Starting date:
January 14, 2016
Posting date:
February 8, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57022

Affected Products

A. TRILOGY 100 VENTILATOR-CANADA

B. TRILOGY 100 VENTILATOR-INTERNATIONAL

C. TRILOGY 202, INTERNATIONAL

D. TRILOGY 202, DOMESTIC

E. TRILOGY 200, CANADA

Reason

Trilogy Ventilators with software versions 13.2.04, 13.2.05, 14.0.00 or 14.1.01 are susceptible to an inadvertent change between prescriptions. This may occur under a particular set of operating conditions and after a specific sequence of key presses without requiring confirmation of the change by the user. Affected devices continue to accurately display the active prescription in the upper left hand corner of the display. As this prescription change was unintended and no confirmation was required, the user may be unaware of this change.

Affected products

A. TRILOGY 100 VENTILATOR-CANADA

Lot or serial number

All

Model or catalog number
  • CA1054096
  • R1054260
Companies
Manufacturer

Respironics Inc.

1001 Murray Ridge Lane

Murraysville

15668

Pennsylvania

UNITED STATES



B. TRILOGY 100 VENTILATOR-INTERNATIONAL

Lot or serial number

All

Model or catalog number

1054096

U1054096

Companies
Manufacturer

Respironics Inc.

1001 Murray Ridge Lane

Murraysville

15668

Pennsylvania

UNITED STATES



C. TRILOGY 202, INTERNATIONAL

Lot or serial number

All

Model or catalog number

1040000

CA1040000

Companies
Manufacturer

Respironics Inc.

1001 Murray Ridge Lane

Murraysville

15668

Pennsylvania

UNITED STATES



D. TRILOGY 202, DOMESTIC

Lot or serial number

All

Model or catalog number

1040007

Companies
Manufacturer

Respironics Inc.

1001 Murray Ridge Lane

Murraysville

15668

Pennsylvania

UNITED STATES



E. TRILOGY 200, CANADA

Lot or serial number

All

Model or catalog number

CA1032800

Companies
Manufacturer

Respironics Inc.

1001 Murray Ridge Lane

Murraysville

15668

Pennsylvania

UNITED STATES