This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TRIDENT CONSTRAINED LINER IMPACTOR TIPS (2016-06-06)

Starting date:
June 6, 2016
Posting date:
June 24, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-59024

Affected products

A. TRIDENT CONSTRAINED LINER IMPACTOR TIPS

Reason

During steam sterilization validation it was discovered that the TRIDENT Constrained Liner Impactor Tips do not meet the required Sterility Assurance Level (SAL) of 10-6. Please note that these instruments are sterilized at the hospital prior to surgery, however, and do not come from Stryker sterile.

Affected products

A. TRIDENT CONSTRAINED LINER IMPACTOR TIPS

Lot or serial number

All lots

Model or catalog number
  • 2199-2022
  • 2199-2028
  • 2199-2032
Companies
Manufacturer

Howmedica Osteonics Corporation

325 Corporate Drive

Mahwah

New Jersey

UNITED STATES