This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Triathlon Tibial Alignment Handle

Starting date:
October 11, 2011
Posting date:
January 9, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-15118

Recalled products

  1. Triathlon Tibial Alignment Handle

Reason

Stryker  received a product experience report (PER) that a Swivel Block was improperly oriented on a Triathlon Tibial Alignment handle. The locator holes on the alignment guide handle on which the Swivel Block is mounted are not in alignment as per print. The Tibial Alignment handle was not machined properly on the 5 Axis machine because the Tibial Alignment handle was not fixtured properly on the machine by the operator. The fixturing procedure involved manually locating every part in a collet, aligning Datum C using a square and locking the part in position for machining of the slot and cross hole. Additionally, the particular feature was not inspected during in-process and/or final inspection at the vendor. Hence, the non-conformance was not detected.

Affected products

A. Triathlon Tibial Alignment Handle

Lot or serial number
  • > 10 Numbers, Contact Manufacturer
Model or catalog number
  • 6541-2-807
  • 6541-2-808
  • I-K2072AHOO
Companies
Manufacturer
Howmedica Osteonics Corporation