Triathlon Tibial Alignment Handle
- Starting date:
- October 11, 2011
- Posting date:
- January 9, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-15118
Recalled products
- Triathlon Tibial Alignment Handle
Reason
Stryker received a product experience report (PER) that a Swivel Block was improperly oriented on a Triathlon Tibial Alignment handle. The locator holes on the alignment guide handle on which the Swivel Block is mounted are not in alignment as per print. The Tibial Alignment handle was not machined properly on the 5 Axis machine because the Tibial Alignment handle was not fixtured properly on the machine by the operator. The fixturing procedure involved manually locating every part in a collet, aligning Datum C using a square and locking the part in position for machining of the slot and cross hole. Additionally, the particular feature was not inspected during in-process and/or final inspection at the vendor. Hence, the non-conformance was not detected.
Affected products
A. Triathlon Tibial Alignment Handle
Lot or serial number
- > 10 Numbers, Contact Manufacturer
Model or catalog number
- 6541-2-807
- 6541-2-808
- I-K2072AHOO
Companies
- Manufacturer
- Howmedica Osteonics Corporation