This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Triage Cardio3 Panel (2014-02-15)

Starting date:
February 15, 2014
Posting date:
March 25, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38589

Recalled Products

Plaque Triage Cardio3

Reason

Alere Triage Cardio 3 lot number W53658B was investigated in response to data observed as part of the routine product performance monitoring program. This lot shows lower than expected results for BNP, CKMB and Troponin I, which in turn could result in an increased rate of false negative results. Testing of external controls would enable your lab to detect if this lot was giving lower than expected results at the time of your testing.

Affected products

Plaque Triage Cardio3

Lot or serial number

97400EU

Model or catalog number

W53658B

Companies
Manufacturer

Alere San Diego, Inc.

9975 Summers Ridge Road

San Diego

92121

California

UNITED STATES