Tri-Staple™ 2.0 Black Intelligent Reload
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Tri-Staple™ 2.0 Black Intelligent Reload |
N2D0195Y |
SIG60AXT |
Issue
Specific lots have the potential for a broken sled vane component. A broken sled vane may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may potentially be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
Medtronic has received one complaint related to this issue causing serious injury, which included a delay in treatment and tissue trauma. No deaths have been reported.
Recall start date: January 25, 2023
Additional information
Details
Covidien Llc
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048
Get notified
Receive emails about new and updated recall and safety alerts.