Trace Elements Normal Range Serum Control (2019-11-22)
- Starting date:
- November 22, 2019
- Posting date:
- December 13, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71869
Last updated: 2019-12-13
Affected Products
Trace Elements Normal Range Serum Control
Reason
Incorrect value on the product IFU for the analyte of Manganese. The value was mistyped as a value of 2.69 and range of 2.29 - 3.09 on the IFU, when it was reported as 0.48 and 0.41 - 0.55 respectively.
Affected products
Trace Elements Normal Range Serum Control
Lot or serial number
C2540
Model or catalog number
66816
Companies
- Manufacturer
-
UTAK LABORATORIES INC.
25020 AVENUE TIBBITTS
VALENCIA
91355
California
UNITED STATES