This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TPHA Screen Test Cells

Starting date:
May 18, 2017
Posting date:
June 12, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63586

Reason

Immucor, Inc. is issuing this medical device recall regarding the stability of the TPHA screen test cells, lot 280006. A review of post-release testing demonstrated this lot of TPHA screen test cells had a decrease in sensitivity for detection of antibodies to treponema pallidum over the product's shelf life.

Affected products

TPHA Screen Test Cells

Lot or serial number

280006

Model or catalog number

0066826

Companies

Manufacturer

Immucor Inc.

3130 Gateway Drive, P.O. Box 5625

Norcross

30091-5625

Georgia

UNITED STATES