Health product recall

Tina-Quant IGG Gen 2 (2018-06-07)

Starting date:
June 7, 2018
Posting date:
June 22, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67086

Affected products

Tina-Quant IGG Gen 2

Reason

Internal investigation has revealed that when using the urine application of the Tina-quant IgG Gen.2 Assay (IgGu2) on Cobas C 311/501/502 analyzers with the reagent lot #24242601, a higher signal levels may cause a misfit of the calibration curve, resulting in over-recovery with a maximal bias of +37% for lower concentration samples.

Affected products

Tina-Quant IGG Gen 2

Lot or serial number

24242601

Model or catalog number

03507432190

Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY