Tina-Quant IGG Gen 2 (2018-06-07)
- Starting date:
- June 7, 2018
- Posting date:
- June 22, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67086
Affected products
Tina-Quant IGG Gen 2
Reason
Internal investigation has revealed that when using the urine application of the Tina-quant IgG Gen.2 Assay (IgGu2) on Cobas C 311/501/502 analyzers with the reagent lot #24242601, a higher signal levels may cause a misfit of the calibration curve, resulting in over-recovery with a maximal bias of +37% for lower concentration samples.
Affected products
Tina-Quant IGG Gen 2
Lot or serial number
24242601
Model or catalog number
03507432190