Health product recall

Tina-Quant IgA Gen2 (IgA-2) (2019-05-27)

Starting date:
May 27, 2019
Posting date:
June 21, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70225


Last updated:

Affected Products

Tina-Quant IgA Gen2 (IgA-2)

Reason

There are potential calibration and quality control material failures for Tina-Quant IgA Gen. 2 assay (cat # 05219205190 lot 368756) when used on the cobas c 701-702 analyzers. This situation could lead to an under-recovery and/or over recovery of IgA in serum/plasma. This situation is only limited to a number of c packs, it can be detected by quality control or calibration failures of the affected reagent c packs.

Affected products

Tina-Quant IgA Gen2 (IgA-2)

Lot or serial number

368756

Model or catalog number

5219205190