This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

THERMOGUARD DISPERSIVE ELECTRODES (2016-05-26)

Starting date:
May 26, 2016
Posting date:
June 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58894

Affected products

  1. THERMOGUARD DISPERSIVE ELECTRODES

Reason

A version of the ConMed ThermoGard Dual Dispersive Electrodes, manufactured between October 1, 2014, and July 5, 2015, may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn. A customer alerted ConMed of this condition on April 6, 2016. These generators use a CQM (contact quality monitoring) technology which may rely on a preset upper limit (alarm limit) of the CQM impedance to detect pad lift. Please contact ConMed for a complete list of compatible CQM electrosurgical generators (esu's).

Affected products

A. THERMOGUARD DISPERSIVE ELECTRODES

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 51-7310
  • 7-382
Companies
Manufacturer
Conmed Corporation
525 French Road
Utica
13502
New York
UNITED STATES