Tempus Pro
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Tempus Pro |
Not applicable. |
00-1026-R |
Tempus Pro |
Not applicable. |
00-1004-R |
Tempus Pro |
Not applicable. |
00-1007-R |
Tempus Pro |
Not applicable. |
00-1024-R |
Issue
The user/operator manual references the term 'invasive pressure', which includes intracranial pressure (ICP) and bladder pressure (BDR) as selectable channel labels.
However, these measurements on the Tempus Pro monitor have not been tested or validated for accuracy. The device is only cleared for invasive blood pressure (IBP) measurements.
Additional information
Details
| Remote Diagnostic Technologies Limited |
| Ascent 1, Farnborough, United Kingdom, GU14 6XW |