Health product recall

TechnegasPlus Generator (2019-06-20)

Starting date:
June 20, 2019
Posting date:
July 15, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70415


Last updated:

Affected Products

TechnegasPlus Generator

Reason

On May 24th, 2019, while performing conducted emissions certification testing on the technegasplus generator to CISPR 11, the TechnegasPlus Generator recorded a measurement above the upper limit  for class a conducted emissions. Specifically, the out-of-specification reading occurred during the 17 second burn cycle on the active line testing while assessing quasi-peak levels.

Affected products

TechnegasPlus Generator

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

25000

Companies
Manufacturer

CYCLOMEDICA AUSTRALIA PTY LTD

UNIT 4, 1 THE CRESCENT

KINGSGROVE

2208

AUSTRALIA