Health product recall

TDWorkstation (2018-02-03)

Starting date:
February 3, 2018
Posting date:
April 3, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66246

Affected products

TDWORKSTATION

Reason

Software anomaly (reference dr35041) with TD-Synergy version v11.31 or later. This anomaly results in consequences for users only when the TD-Synergy system is connected to another system and uses one of these two communication methods. Transmission of results in HL7 format from TD-Synergy to a host system including results of date type. Sharing of TD-Synergy data with a reference laboratory including complementary parameters of date type.

This anomaly affects only results of date type or complementary parameters of date type.

Nature of the defect (real or potential):

  • Depending on the date format used, the system may invert the date’s day and month. For example, the system can transmit a date as “February 3, 2015” instead of “March 2, 2015”
  • The system transmits results of date type or review requests with complementary parameters of date type in HL7 format.

The communication module’s service window uses a local account system that has a regional parameter corresponding to common wording that uses a date format in which the day precedes the month (e.g., “day/month/year” format such as “dd/mm/yy” or “d/m/yy”).
Or
The communication module’s service window uses a specific user account, and this user uses a date format in which the day precedes the month (e.g., “day/month/year” format such as “dd/mm/yy” or “d/m/yy”).

The defect does not occur if the date format used is “month/day/year” (mm/dd/yyyy).

Affected products

TDWORKSTATION

Lot or serial number
  • DR35041
  • DR35172
Model or catalog number

TD-WST-01