Health product recall

TCN™ and CSK Reusable RF electrodes

Last updated

Summary

Product
TCN™ and CSK Reusable RF electrodes
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

TCN™, Nitinol

All lots.

TCN-15-3M

TCN™, Nitinol

All lots.

TCN-10-3M

TCN™, Nitinol

All lots.

TCN-5-3M

TCN™, Nitinol

All lots.

TCN-20-3M

TCN™ Electrode

All lots.

TCN-20

TCN™ Electrode

All lots.

TCN-15

TCN™ Electrode

All lots.

TCN-5

TCN™ Electrode

All lots.

TCN-10

CSK Electrode

All lots.

CSK-TC10

CSK Electrode

All lots.

CSK-TC15-3M

CSK Electrode

All lots.

CSK-TC20

CSK Electrode

All lots.

CSK-TC5-3M

CSK Electrode

All lots.

CSK-TC5

CSK Electrode

All lots.

CSK-TC10-3M

CSK Electrode

All lots.

CSK-TC15

Issue

Boston Scientific has determined that the TCN™, and CSK reusable RF electrodes may not consistently meet the expected performance levels throughout the specified 20 sterilization cycles for re-use.  This determination was based on recent review of internal laboratory testing.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies

Boston Scientific Neuromodulation Corporation

25155 Rye Canyon Loop, Valencia, California, United States, 91355

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82046

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe