This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
TCM4 Series Base Unit
- Starting date:
- November 27, 2012
- Posting date:
- January 7, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-16531
Recalled Products
A. TCM4 Series Base Unit
Reason
As reported to Radiometer Medical APS, there is a potential risk of receiving an electrical shock under certain conditions. TCM4 series base unit does not comply with IEC 60601-1, 2nd edition, subclause 15(b).
Affected products
A. TCM4 Series Base Unit
Lot or serial number
000-154, 391-876, 391-878, 391-880, 910-246, Serial Numbers below R0175N009
Model or catalog number
000-154, 391-876, 391-878, 391-880, Power Supply 910-246
Companies
- Manufacturer
-
Radiometer Medical APS