Health product recall

TAQPATH COVID-19 RT-PCR KIT & APPLIED BIOSYSTEMS COVID-19 INTERPRETIVE SOFTWARE (2020-10-09)

Starting date:
October 9, 2020
Posting date:
October 23, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74161



Last updated:
2020-10-22

Affected Products

A. TAQPATH COVID-19 COMBO KIT

B. TAQPATH COVID-19 RT-PCR KIT

C. APPLIED BIOSYSTEMS COVID-19 INTERPRETIVE SOFTWARE

D. APPLIED BIOSYSTEMS COVID-19 INTERPRETIVE SOFTWARE

Reason

The first issue is related to the MS2 assay which detects the Internal Positive Control (IPC). We have made a software update to the MS2 settings to reduce the risk of invalid, potential false negatives, or inconclusive tests and to decrease the retest burden on customers. Analysis settings for other targets have not been changed.

The second issue is related to instructions for vortexing and centrifugation of RT-PCR reaction plates. Failure to adequately vortex and centrifuge reaction plates during processing may result in anomalous fluorescence and/or an irregular baseline at the beginning of cycling, interfering with analysis and potentially leading to inconclusive or false positive results.

Affected products

A. TAQPATH COVID-19 COMBO KIT

Lot or serial number

Not applicable.

Model or catalog number

A47814

A47817

Companies
Manufacturer

Thermo Fisher Scientific

168 Third Avenue, Waltham

Massachusetts

02451

UNITED STATES


B. TAQPATH COVID-19 RT-PCR KIT

Lot or serial number

Not applicable.

Model or catalog number

A47817

Companies
Manufacturer

Thermo Fisher Scientific

168 Third Avenue, Waltham

Massachusetts

02451

UNITED STATES


C. APPLIED BIOSYSTEMS COVID-19 INTERPRETIVE SOFTWARE

Lot or serial number

Not applicable.

Model or catalog number

COVID-19 SOFTWARE V1.2

Companies
Manufacturer

Thermo Fisher Scientific

168 Third Avenue, Waltham

Massachusetts

02451

UNITED STATES


D. APPLIED BIOSYSTEMS COVID-19 INTERPRETIVE SOFTWARE

Lot or serial number

Not applicable.

Model or catalog number

COVID-19 SOFTWARE V2.0

Companies
Manufacturer

Thermo Fisher Scientific

168 Third Avenue, Waltham

Massachusetts

02451

UNITED STATES