This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TANDEM INTL BIPOLAR SHELL/LINER

Starting date:
September 6, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60360

Affected products

Reason

The affected devices were manufactured with an out of specification retainer groove location. In the event the failure occurs, it would be obvious to the user in that the devices would not provide an audible indicator that the poly lock has engaged. In the event the construct is tight, the user should notice the limited range of motion. However, in the worst case situation, if the device gives a false indicator of locking, the user would not recognize the failure.

Affected products

TANDEM INTL BIPOLAR SHELL/LINER

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

More than 10 numbers, contact manufacturer

Companies
Manufacturer

Smith & Nephew Inc.

1450 Brooks Road

Memphis

38116

Tennessee

UNITED STATES