This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

System DI-60 Main Unit and Remote Review Software (2017-05-05)

Starting date:
May 5, 2017
Posting date:
June 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63536

Affected Products

A. System Di-60 Main Unit

B. System Di-60 Remote Review Software

Reason

The DI-60 automated digital cell morphology analyzer shuts down the database and stops processing slides after 100 consecutive slides are run on one day.  The following error message displays, "No connection with the database."  This is a software defect and impacts only the DI-60 and DI-60 with system DI-60 remote review software.

Affected products

A. System Di-60 Main Unit

Lot or serial number

60060-60769 with 6.0.1SOFTWARE

Model or catalog number

CC286297

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN



B. System Di-60 Remote Review Software

Lot or serial number

60060-60769 with 6.0.1SOFTWARE

Model or catalog number

BC188562

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN