This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

System DI-60 Main Unit and Remore Review Software (2017-07-27)

Starting date:
July 27, 2017
Posting date:
September 5, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64384

Affected Products

  1. System Di-60 Main Unit
  2. System Di-60 Remote Review Software

Reason

Sysmex America, Inc. has been made aware of a software defect on versions 5.0.1 and 6.0.1 running on the Sysmex DI-60 that may cause the operator to misidentify a slide flagged with a barcode error. If the Sysmex DI-60 processes a slide with an unreadable barcode, the barcode reader will take a picture of the barcode. The software will assign the slide a temporary id beginning with err in the format of erryyyymmddhhmmss (yyyy=year, mm=month, dd=day, hh=hour, mm=minute, ss=second). The image of the barcode will be stored in the database and displayed in the order data screen within the software so the customer can later identify the slide and reassign it with the correct sample id for the slide. If a subsequent slide is manually loaded on the Sysmex DI-60 and has a barcode with a err prefix, the slide will be linked with the image of the barcode from the prior slide the Sysmex DI-60 found unreadable. This can give the false impression the Sysmex DI-60 barcode reader could not read the barcode of the subsequent slide with a barcode with the err prefix. If the user is not careful there is a risk the user will assign the slide with the err prefix printed on it with an incorrect order id via the image captured from a different slide. The SP-10 will automatically auto-assign the sample id when unable to read the barcode on the tube in the format of err050000000001 and auto-increment. These slides are not automatically routed to the Sysmex DI-60.

Affected products

A. System Di-60 Main Unit

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

CC286297

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN



B. System Di-60 Remote Review Software

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

BC188562

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN