Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71861

Last updated:

Affected Products

  1. System Control N+P, STA - Coag Control N+P
  2. Unicalibrator, STA

Reason

Following a customer complaint, Stago has investigated and confirmed that a positive bias exists on the STA - Unicalibrator levels with STA - deficient IX and STA - deficient VIII when compared to the international Standard (IS) for factor VIII or IX. This bias is also observed on several external quality assessment results.

Affected products

A. System Control N+P, STA - Coag Control N+P

Lot or serial number

253145
253416
253655*
254040*
254206
254401
254573
254760

Model or catalog number

678

Companies
Manufacturer
Diagnostica Stago SAS
9, Rue Des Freres Chausson
Asnieres
92600
FRANCE

B. Unicalibrator, STA

Lot or serial number

253180
253964
254721
255108

Model or catalog number

675

Companies
Manufacturer
Diagnostica Stago SAS
9, Rue Des Freres Chausson
Asnieres
92600
FRANCE