Sysmex XN-L Series (2017-12-01)
- Starting date:
- December 1, 2017
- Posting date:
- January 8, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65566
Affected Products
Sysmex XN-L Series
Reason
After the change to standard time from Daylight Savings Time, the analyzer's time may not be consistent with local time. The analyzer date may also not reflect the current date. The discrepancy is due to a time difference between the windows operating system and the XN-L program time. After changing the time and date in the analyzer's software and cycling power to the analyzer, the date and time may not be correct when the analyzer restarts. Patient and Quality Control (QC) results are not impacted. However, the analysis date and time displayed, printed and transmitted with the results may not reflect the actual date and time the sample was run.
Affected products
Sysmex XN-L Series
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
XN-450
XN-550
Companies
- Manufacturer
-
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN