Sysmex XN-10 Hematology Analyzer (2019-01-14)
- Starting date:
- January 14, 2019
- Posting date:
- January 31, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68962
Affected Products
Sysmex XN-10 Hematology Analyzer
Reason
Sysmex has identified a defect in the shielding used for the flow cytometry module (FCM) detector wiring harness in certain XN-10 analyzers. Improper shielding of the FCM detector may result in abnormal scattergram patterns and increased flagging. Numeric results are not affected.
Affected products
Sysmex XN-10 Hematology Analyzer
Lot or serial number
35739
36053
14166
Model or catalog number
XN-10