This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Sysmex XE Analyzers (2016-03-30)

Starting date:
March 30, 2016
Posting date:
April 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58058

Affected Products

A. ANALYSEUR SYSMEX XE-2100D

B. ANALYSEUR SYSMEX XE-2100L

C. ANALYSEUR SYSMEX XE-5000

D. ANALYSEUR SYSMEX XE-2100

Reason

Some packages of the replacement part, piercer set #1 (part # 971-0583-5), used on Xe-Series analyzers have an incorrect piercer inside the box.

The piercer set #1 is a consumable that is periodically replaced by the operator after a number of cycles on the Xe-Series automated hematology analyzers. A small number of boxes labeled as containing a piercer set #1 contain the incorrect piercer that is used on another Sysmex analyzer model.

Affected products

A. ANALYSEUR SYSMEX XE-2100D

Lot or serial number

Not applicable

Model or catalog number

XE-2100D

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

JAPAN



B. ANALYSEUR SYSMEX XE-2100L

Lot or serial number

Not applicable

Model or catalog number

XE-2100L

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

JAPAN



C. ANALYSEUR SYSMEX XE-5000

Lot or serial number

Not applicable

Model or catalog number

XE-5000

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

JAPAN



D. ANALYSEUR SYSMEX XE-2100

Lot or serial number

Not applicable

Model or catalog number

XE-2100

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

JAPAN