Health product recall

Sysmex Blood Coagulation Analyzer CS-2000i

Starting date:
May 14, 2013
Posting date:
June 17, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-34207

Recalled products

  1. Sysmex Blood Coagulation Analyzer CS-2000i

Reason

Single factor applications at the CS-2000i systems might show increased values if a low sample, at the lower end of the measuring range, is followed by a high sample, at the upper measuring range. The extent of carry over and the clinical impact are different depending on the factor assay

Affected products

A. Sysmex Blood Coagulation Analyzer CS-2000i

Lot or serial number

Not Applicable

Model or catalog number
  • 10488584
Companies