Syngo.Plaza (2019-11-14)
- Starting date:
- November 14, 2019
- Posting date:
- November 29, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71748
Last updated: 2019-12-02
Affected Products
Syngo.Plaza
Reason
Software update provides feature enhancements and the resolves various defects including the below. These fixes do not affect patient safety.
- In a cluster environment the studies getting mixed up and patient list information does not match the image dicom info
- Failure in de-archive from dicom lTA (syngo.Share) if study has multiple character sets including non-extendable character set for e.g. ISO_IR 100 and ISO_IR 192
- In dicom communication, normal string copy is now replaced by secured string copy throughout export workflow and dicom libraries
- Drag & drop function across studies from patient jacket was not allowed
Affected products
Syngo.Plaza
Lot or serial number
Not applicable.
Model or catalog number
11293724
Companies
- Manufacturer
-
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY