SYNCHRON SYSTEMS CHOLESTEROL (CHOL) REAGENT
Brand(s)
Summary
Contact the manufacturer if you require additional information
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
SYNCHRON SYSTEMS CHOLESTEROL (CHOL) REAGENT |
M106450 |
467825 |
Issue
Beckman coulter has identified that Synchron Systems Cholesterol Reagent lot M106450 may fail with out of calibration range low (OCR low) and/or back to back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Recall start date: 2021-11-08 |
Additional information
Details
BECKMAN COULTER CANADA L.P.
7075 FINANCIAL DRIVE, MISSISSAUGA
ON, CANADA
L5N 6V8