This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Synchron Systems Aqua Cal 1

Starting date:
May 15, 2017
Posting date:
June 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63542

Reason

Beckman Coulter has confirmed a negative shift of 5 mmol/L for sodium recovery in Synchron Control level 3 when calibrated with Aqua CAL 1 Lot 72645074. A linearity study using Maine linearity standards also indicated a negative shift of up to 8 mmol/L at values between 150 and 180 mmol/L.

Affected products

Synchron Systems Aqua Cal 1

Lot or serial number

72645074

Model or catalog number

471288

Companies

Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES