Health product recall

Symbia T series Gamma Camera/CT Scanners (2019-07-12)

Starting date:
July 12, 2019
Posting date:
July 26, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70527

Last updated: 2019-07-26

Affected Products

Symbia T series Gamma Camera/CT Scanners

Reason

A defect in the broad quantification option of the Symbia product software version vb20a may cause the system not to retain:

  1. The phantom volume entered by the user during volume sensitivity calibration.
  2. The residual dose entered by the user during volume sensitivity calibration or acquisition.

The failure only occurs when broad quantification is purchased without other xSPECT Quant options. The software update vb20b service pack will eliminate the issue, correct other minor software problems and improve performance.

Affected products

Symbia T series Gamma Camera/CT Scanners

Lot or serial number

Not applicable.

Model or catalog number

10275007
10275008
10275009

Companies
Manufacturer
Siemens Medical Solutions USA, Inc. Molecular Imaging
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES