Health product recall

Swan Ganz Thermodilution Catheter (2018-12-17)

Starting date:
December 17, 2018
Posting date:
January 23, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68830

Affected products

Swan Ganz Thermodilution Catheter

Reason

This non-conformance involves Swan Ganz catheters with incorrect lumen assembly, which causes reversal of the lumens.  If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.  It is possible for unintended treatment due to inaccurate values if the reversal in waveforms is not noticed.

Affected products

Swan Ganz Thermodilution Catheter

Lot or serial number
  • 61176367
  • 61176369
  • 61321241
Model or catalog number
  • 131F7
  • 131F7P
  • 151F7
Companies
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
UNITED STATES