Stryker ProCuity Beds
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Stryker ProCuity Beds |
More than 10 numbers, contact manufacturer. |
3009PX-LE-400 |
Issue
Stryker has identified that the manufacturing records for certain ProCuity beds are missing test values for electrical safety tests as required by IEC 62353:2014. As a result of the missing test data, the impacted units are not compliant to the above standard. Therefore, Stryker is initiating this voluntary recall to perform the necessary testing and documentation in the field.
Recall start date : April 29, 2024
Additional information
Details
Stryker Medical
3800 E Centre Avenue, Portage, Michigan, United States, 49002