Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Stryker ProCuity Beds

More than 10 numbers, contact manufacturer.

3009PX-LE-400
300900000000
3009QC-P1
3009PX-LEX-400

Issue

Stryker has identified that the manufacturing records for certain ProCuity beds are missing test values for electrical safety tests as required by IEC 62353:2014. As a result of the missing test data, the impacted units are not compliant to the above standard. Therefore, Stryker is initiating this voluntary recall to perform the necessary testing and documentation in the field.

Recall start date : April 29, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Stryker Medical

3800 E Centre Avenue, Portage, Michigan, United States, 49002

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75497