Stratus CS Acute Care System Cardiac troponin-i TestPak (2019-06-04)
- Starting date:
- June 4, 2019
- Posting date:
- June 21, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70257
Last updated: 2019-06-21
Affected Products
Stratus CS Acute Care System Cardiac troponin-i TestPak
Reason
Increased rate of false positive cTnI results beginning 2 months prior to the end of the product's shelf life due to a positive bias and increased imprecision. Increased rate of random non-repeatable false positive cTnI results at any time in the product's shelf life.
Affected products
Stratus CS Acute Care System Cardiac troponin-i TestPak
Lot or serial number
238330002
239007002
239014002
239028002
239035002
239056002
239063002
239070002
239077002
239105002
239112002
239119002
Model or catalog number
CCTNI
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, P.O. Box 6101
Newark
19714
Delaware
UNITED STATES