Health product recall

StraightShot Microdebrider System XPS Blades (2018-05-30)

Starting date:
May 30, 2018
Posting date:
June 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67122

Affected products

StraightShot Microdebrider System XPS Blades

Reason

This voluntary recall is being conducted due to potential for irrigation port blockage. The irrigation pin used to manufacture (molding) the hub outer tube was incorrectly placed in the mold which resulted in blockage of the irrigation lumen. Medtronic Puerto Rico Operations Company (MPROC) presented an issue in which the 66d1204 hub outer tube w/barb 3.5mm molded hub manufactured at Medtronic ENT Jacksonville had one cavity with a blocked irrigation port.

Affected products

StraightShot Microdebrider System XPS Blades

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 1883504
  • 1883507
  • 1883514
  • 1883516
  • 1883523
  • 1883524
  • 1883525