StraightShot Microdebrider System XPS Blades (2018-05-30)
- Starting date:
- May 30, 2018
- Posting date:
- June 29, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67122
Affected products
StraightShot Microdebrider System XPS Blades
Reason
This voluntary recall is being conducted due to potential for irrigation port blockage. The irrigation pin used to manufacture (molding) the hub outer tube was incorrectly placed in the mold which resulted in blockage of the irrigation lumen. Medtronic Puerto Rico Operations Company (MPROC) presented an issue in which the 66d1204 hub outer tube w/barb 3.5mm molded hub manufactured at Medtronic ENT Jacksonville had one cavity with a blocked irrigation port.
Affected products
StraightShot Microdebrider System XPS Blades
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 1883504
- 1883507
- 1883514
- 1883516
- 1883523
- 1883524
- 1883525
Companies
- Manufacturer
-
Medtronic Xomed Inc.
6743 Southpoint Drive North
Jacksonville
32216
Florida
UNITED STATES